CO₂ Laser Tube Antifreeze Usage Instructions
During winter or when the ambient temperature drops below 0°C, antifreeze should be added to the cooling water to prevent the water inside the CO₂ laser tube (hereinafter referred to as the laser tube) from freezing. Alternatively, when the machine is shut down, the water inside the laser tube should be completely drained.
When using antifreeze, please observe the following requirements:
1. Antifreeze Selection
Choose an antifreeze with the following properties:
Good fluidity
Strong freeze protection
Low volatility
Low corrosiveness
Good heat dissipation performance
The freezing point of the antifreeze should be at least 5°C lower than the minimum ambient temperature expected during operation to reduce the risk of the laser tube cracking due to freezing in extremely cold weather.
In general, higher antifreeze concentrations result in poorer fluidity and heat dissipation. Customers should select an appropriate concentration according to actual operating conditions.
Always handle antifreeze safely and follow the manufacturer's instructions.
Commonly available antifreeze is generally made by mixing ethylene glycol with water. The approximate relationship between ethylene glycol concentration and freezing point is shown below.
2. Ethylene Glycol Concentration and Freezing Point
| No. | Ethylene Glycol Concentration | Freezing Point | No. | Ethylene Glycol Concentration | Freezing Point | No. | Ethylene Glycol Concentration | Freezing Point |
|---|---|---|---|---|---|---|---|---|
| 1 | 20% | -9°C | 4 | 40% | -25°C | 7 | 51% | -40°C |
| 2 | 28% | -15°C | 5 | 43% | -31°C | 8 | 54% | -45°C |
| 3 | 34% | -20°C | 6 | 46% | -35°C | 9 | 60% | -50°C |
Note: Freezing points are approximate reference values. Actual performance may vary depending on the antifreeze formulation and product specifications.
3. Check Laser Performance After Adding Antifreeze
Before and after replacing the cooling water with antifreeze, carefully compare the laser tube's cutting speed and beam quality.
If any of the following problems occur after adding antifreeze, it may be affecting the laser tube's output power or beam quality:
Reduced cutting speed
Unusual noise when cutting acrylic
Noticeable deterioration in laser spot quality
If these symptoms occur, stop using the current antifreeze and replace it with another suitable product after checking its compatibility with the laser tube and cooling system.
4. Replacing Antifreeze with Purified Water
When the ambient temperature rises above 0°C, replace the antifreeze with purified water in a timely manner.
Before refilling, completely drain the antifreeze from the following components:
Water chiller
CO₂ laser tube